Acopian P01.5HP20 High Voltage AC DC Power Supplies. Output Range kVdc 0-1.5 | Output Current mA 20 | Rack Mounting Regulated Simple Type: Power Supply

Ideal for laboratory and instrumentation applications, these rack mounting supplies feature calibrated ten-turn controls (with locking vernier dials) for precisely setting voltage and current. Voltmeter, ammeter and handles are standard.

Download the datasheet (PDF)

For Resellers and High Volume Orders:
Please request a quote to obtain preferred pricing.

Features

  • Front panel adjustment controls with locking vernier dials
  • Input protection
  • Output programming
  • Output protection
  • Remote monitoring of output current
  • Remote monitoring of output voltage
  • Standard Voltmeter and ammeter

Specifications

    General
  • Ambient operating temperature: –10 to +60°C | No derating required
  • Current MonitorTerminal: Permits remote monitoring of output current, at mV/mA ratio shown. Accuracy is 2% of maximum rated output current
  • Humidity: Maximum of 90% relative, non-condensing
  • InhibitTerminal: Grounding inhibits output
  • Input Current: 30 watt output ratings: 0.6A | 60 watt output ratings: 1.2A
  • Input Protection: “Soft start” circuit minimizes start-up power stresses
  • Input, Control and Monitoring: Screw terminals
  • InputVoltage: 105-125 VAC, 50-400 Hz, single phase.
  • Metering: Voltmeter and ammeter are standard. Accuracy is 2% of full scale
  • Output: High voltage connector (Type varies with model number)
  • Output Controls: Voltage and current may be controlled by means of two 10-turn front panel adjustments with locking vernier dials. Control linearity is 1% of full rated output. Calibration accuracy is 1% of rated output plus 1% of setting. (Remotely located 1000 ohm potentiometers may alternately be used for output control)
  • Output Polarity: Positive output is standard. For negative output, change first letter of model number from P to N
  • Output Programming: Output voltage and current may be programmed from 0 to full rating by means of control voltage inputs of 0 to +5.1 Vdc, ±2%
  • Output protection: Current regulation circuit protects power supply from short circuits, overload, and arcing
  • Regulation (constant current operation): (Line) ±0.1% | (Load:)±0.1% plus 50 µA
  • Regulation (constant voltage operation): (Line) ±0.05% | (Load) ±0.05%
  • Response Time: Less than 5 mS for 100 µA load step change
  • Ripple: 0.05%, peak-to-peak
  • STABILITY: 0.05% over eight hours, after 30 minute warmup
  • StorageTemperature: –20 to +85°C
  • Temperature Coefficient: 200 PPM/°C equals 0.02%/°C (Typical)
  • Voltage Monitor Terminal: Permits remote monitoring of output voltage, stepped down by ratio shown.Accuracy is 2% of maximum rated output voltage

 